Newtronic Photostability Chambers are
specifically designed to meet the requirements for near UV and visual light
testing as outlined in ICH Q1B Guidelines, Option 2. These chambers are
ideal for forced degradation testing and confirmatory studies, which help
identify precautions needed in manufacturing, formulation, and whether
light-resistant packaging is necessary. With high-intensity fluorescent lamps,
uniform light distribution, and precise temperature and humidity control, the
chambers ensure optimal conditions for photostability testing. The
user-friendly PLC-based control system, mobile alerts, and advanced features
such as auto-switch-off of UV lights upon door opening provide seamless
operation. Newtronic chambers also feature specular reflective surfaces that
ensure proper spectral power distribution (SPD) for reliable testing results.
Key
Features:
Ethernet Communication:
High-speed
data transfer surpassing serial communication (RS-485).
Global
standard communication protocol for remote operation via LAN without
requiring a separate PC interface.
Mobile Alert System:
Real-time
alerts sent directly to registered devices.
Transmitted
message includes equipment name, alarm time, alarm type, and limits.
ICDAS Software:
21 CFR
Part 11 ready software.
Capabilities
include e-signatures, automatic daily backups, alarm logs, audit trails,
MKT calculations, graphical data trending, and PDF data export.
Comprehensive Alarm System:
Alerts
for temperature & humidity deviations, door opening, water supply
failure, component failure, and mains failure.
Safety Features:
Protection
from temperature overshoot, humidity overshoot, low water levels, and
over-current with an MCB cut-off.
Standby
sensor upon primary system failure.
Emergency
bell switch inside the chamber.
PLC-Based Control System:
Complete
control via PLC.
No
data loss: Stores up to 1000 readings at 30-minute intervals for up to 20
days.
Direct
email notifications from the PLC for critical updates.
Touch Screen Display:
Easy-to-use
4” color touch screen for all operations.
Features
include PLC IP address setting, component working hours, standby system
changeover, and password protection for calibration.
Documentation & Training:
Full
suite of documentation: IQ, OQ, PQ, calibration certificates, operations
& maintenance manuals, GA & electrical drawings, material test
certificates, and user training.
Write a review