Ask a Question About This Product
LabSen® 241-180 Micro pH Electrode for NMR Tubes
- Manufacturer: Apera Instruments
- Item Number: AI3161
Price excludes taxes and shipping charges.
295.00€
The LabSen® 241-180 micro pH electrode from Apera Instruments is specifically designed for precise pH measurement in NMR tubes and other thin vessels. Manufactured with Swiss sensor technology, this electrode features a 180mm cylindrical micro glass membrane, low membrane resistance, and a long-life reference system for fast and stable measurements, even in high-temperature environments (up to 100°C).
Key Features:
- Swiss Sensor Technology – Advanced high-precision design for reliable pH measurements in small-volume samples.
- Cylindrical Micro Glass Membrane –
- Length: 180mm
- Diameter: 3mm
- Low Membrane Resistance (<450MΩ) – Ensures fast response times and high measurement stability.
- High-Temperature Compatibility – Usable in temperatures up to 100°C.
- Long-Life Reference System –
- Consists of: Glass tube, AgCl, and reference silver wire.
- Cotton-Filled Glass Tube – Prevents reaction between AgCl and electrolyte, ensuring stability across temperature changes and extending electrode lifespan.
- Small Volume Measurement – Ideal for NMR tubes and thin vessels, requiring a minimum sample volume of 50μl.
Compatibility:
- BNC Connector – Compatible with all Apera Instruments pH meters using BNC connectors, including:
- PH850, PC850, PH8500, PC8500, PH700, PH910, PC910, PH950, PC950, PH800, PC800, PH820, and PC820.
- Temperature Compensation – Requires a separate temperature sensor (e.g., Apera Instruments MP500) for automatic temperature compensation or manual temperature input.
Warranty:
- 6-Month Manufacturer's Warranty ensuring long-term reliability for NMR tube pH measurements.
Technical Specifications | |
Cable | Ø3mm×1m |
Diaphragm | Ceramics |
Electrode Dimension | Electrode (Ø12×60) mm Measuring tip (Ø3×180) mm |
Electrolyte | 3M KCl |
Glass Membrane Type | S |
Measuring Range | (0 ~ 14) pH |
Membrane resistance | <450 MΩ (@25°C) |
Microvolume | ≥50μl |
Recommended operating temperature | (0 ~100)°C |
Reference system | Long Life |
Shank material | Lead free glass |
Storage solution | 3M KCl |